產品資料

9245 鈉表

如果您對該產品感興趣的話,可以
產品名稱: 9245 鈉表
產品型號: polymetron
產品展商: Polymetron
會員價格: 0.00 元
產品文檔: 無相關文檔

簡單介紹

polymetron 9245 Single Channel On-Line Sodium Analyzer


9245 鈉表  的詳細介紹
Details
 
9245 Single Channel On-Line Sodium Analyzer (0 to 10,000 ppb) with Enclosure, Universal

9245 Sodium Analyzer

Analyzer has automatic reactivation ensures optimum response time and performance; no acid etching required! 100 days between reagent changes. Includes US fittings kit, 3M KCl, 10 ppm sodium standard, Diisopropylamine (DIPA), and Sodium Nitrate. Available in US and Canada only.

  • Automatic Electrode Reactivation for Optimum Operation and Response Time
  • Easy to Install, Operate, Calibrate, and Maintain
  • Adaptable to a Variety of Conditions
  • Lower Detection Limit of 0.01 ppb
  • 4-20 mA Outputs and Digital Communication Options
  • Automatic Electrode Reactivation for Optimum Operation and Response Time

    To maintain optimum response time, even in systems of continuous low sodium concentration, the analyzer automatically reactivates the electrode using non-hazardous chemicals.

    Easy to Install, Operate, Calibrate, and Maintain

    The 9245 Sodium Analyzer installs easily; it requires only power and sample connections. Clear step-by-step instructions are given by the analyzer to simplify operations such as instrument start-up, long-term stand-by, and reagent replenishment. Low ppb sodium solutions for calibration can be difficult to prepare so the 9245 analyzer is equipped with a fully automated calibration system that uses 10 ppm samples and follows a multiple step calibration cycle to accurately calibrate the analyzer. Normal maintenance consists of replenishment of the calibration solution reagents and electrolyte, typically every 100 days, and annual replacement of the reagent tube.

    Adaptable to a Variety of Conditions

    The analyzer is fitted with a quick-loop valve, a constant head vessel, and minimum sample flow detection to allow for variations in sample pressure, flow, and distance from the analyzer to the fresh-sample point. Grab sample can be obtained for manual calibration, accuracy verification, and measurement of one-off process samples after which, the unit returns automatically to on-line monitoring.

    Lower Detection Limit of 0.01 ppb

    With a detection limit of 0.01 ppb and a range of 0 to 10,000 ppb, the 9245 Sodium Analyzer is ideally suited for monitoring sodium in demineralized water, boiler feed, condensate, and all parts of the steam/water cycle.

    4-20 mA Outputs and Digital Communication Options

    Four sets of isolated analog outputs can be configured in 4-20 mA with three outputs assigned to sample concentration or temperature readings and the fourth reporting electrode live signal, including calibration and grab sample. Additional digital communication is available with JBUS/MODBUS or Profibus DP.
    9245 sodium analyzer, us fittings kit, 3M KCl solution, 10ppm sodium standard, DIPA reagent, Sodium nitrate solution, and Hach manual.

     Specifications

     

    Accuracy:± 0.1 ppb or ±5% of reading, whichever is greater
    Accuracy 2:Cationic application: ± 2 ppb or 5 % of reading, whichever is greater
    Ambient Temperature:5 to 50 °C (41 to 122 °F)
    Communication Capabilities:Standard (4x 4-20 mA outputs)
    Lower Limit of Detection (LOD):0.01 ppb
    Measurement Range 2:0 to 200 ppm with Cation-kit option
    Mounting:Wall mount enclosure
    Panel Dimensions:x x
    Range:0 to 10000 ppb freely programmable
    Repeatability:< 0.02 ppb or 1.5 % of reading, whichever is greater, within 10ºC variation

     

    Response Time:1 cycle, minimum 10 minutes (t=90%)